SMITHS MEDICAL ASD, INC. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number CA24E0/4K2/004JP |
Device Problem
Connection Problem (2900)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/19/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that the connection between the product and anesthesia was not attached.No patient injury was reported.
|
|
Manufacturer Narrative
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms#(b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Unit returned was received inside of a plastic bag which is not the original package.Sample can't be submitted for leak test inspection because the connectors are necessary to fit in the leak tester equipment.Additionally, process assembly inspect the pieces at 100%, they can detect if the product is without connector.Failure reported could not be reproduced on the process.The root cause cannot be confirmed, the most probable root cause occurred after the product left the manufacturing facilities and the sample returned was already open it by the customer.No corrective actions were taken since the investigation did not reveal that the defect was caused during the manufacturing process.No problems or issues were identified during this device history record review.D5: operator of device is unknown.G5: premarket (510k) number is unknown.
|
|