Model Number DTBA1D1
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/23/2022
Event Type  malfunction  
Event Description
The device was returned to the manufacturer after being explanted due to being at normal elective replacement indicator (eri).The device subsequently tested out of specification during manufacturer¿s analysis.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.The returned device indicated shorting/low impedance in the battery.Hybrid analysis revealed that the device loss of telemetry and all functions were due to a severely depleted battery.The device was fully functional and operated under normal current drain when powered with an external supply.There was no evidence of a high current drain condition, and no hybrid anomalies were observed.Battery analysis revealed that the premature battery depletion was the result of a cathode particle short.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA¿ XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key16852863
MDR Text Key314370224
Report Number3004209178-2023-05775
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00643169530140
UDI-Public00643169530140
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/02/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date12/14/2017
Device Model NumberDTBA1D1
Device Catalogue NumberDTBA1D1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/30/2022
Date Manufacturer Received04/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/20/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
5076-52 LEAD, 419488 LEAD, 693565 LEAD.
Patient Age55 YR
Patient SexMale
-
-