SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX ANESTHESIA BREATHING CIRCUIT; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C1034J |
Device Problem
Material Twisted/Bent (2981)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
B3: month and year of event have been provided, day is unknown.D4: udi number is unknown, no information has been provided to date.G5: 510k is blank device is exempt.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
|
Event Description
|
It was reported that before opening the package, twisting of the inner tube of the breathing circuit was found.No adverse patient effects were reported by the customer.It was reported that no further information is available.
|
|
Manufacturer Narrative
|
Regulatory.Responses@icumed.Com.One device sample was received without the original package.Per visual inspection, it was confirmed damaged failure mode.A leak test was performed, and the result of the corrugated tubing assembly test was acceptable because there was no air leak.No further device analysis was performed.The root cause could not be determined however it could have been caused by overhandling of the product.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.The manufacturer will continue monitoring this failure condition in this product for threshold or escalation.
|
|