(b)(4).The actual sample was not returned; however, the customer provided two photos and a video for analysis.The complaint of luer hub/fitting has crack was able to be confirmed by the photos and video.The two photos revealed a crack that runs the length of the blue luer hub.The video revealed the same crack in use which resulted in leaking.A complete visual inspection could not be performed as no sample was returned for analysis.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
|