• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC ARROW CANNON II PLUS REPLACEMENT HUB SET; KIT, REPAIR, CATHETER, HEMODI

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC ARROW CANNON II PLUS REPLACEMENT HUB SET; KIT, REPAIR, CATHETER, HEMODI Back to Search Results
Catalog Number CAR-02800
Device Problems Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/05/2023
Event Type  malfunction  
Event Description
It was reported during dialysis blood was coming through the blue port hub.On inspection it had a large longitudinal crack from where cap fastens to the bulb area.The hub was changed with a hub repair kit.No patient harm reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported during dialysis blood was coming through the blue port hub.On inspection it had a large longitudinal crack from where cap fastens to the bulb area.The hub was changed with a hub repair kit.No patient harm reported.The patient's condition is reported as fine.
 
Manufacturer Narrative
(b)(4).The actual sample was not returned; however, the customer provided two photos and a video for analysis.The complaint of luer hub/fitting has crack was able to be confirmed by the photos and video.The two photos revealed a crack that runs the length of the blue luer hub.The video revealed the same crack in use which resulted in leaking.A complete visual inspection could not be performed as no sample was returned for analysis.The customer did not provide a lot number; therefore, a device history record review was performed based upon a lot number from the sales history data of the customer.No relevant findings were identified.The instructions for use (ifu) provided with this kit warns the user, "repeated over tightening of bloodlines, syringes and caps will reduce connector life and could lead to potential connector failure." without the device to evaluate, the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARROW CANNON II PLUS REPLACEMENT HUB SET
Type of Device
KIT, REPAIR, CATHETER, HEMODI
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL DE MEXICO S.A. DE C.V.
ave. washington 3701
colonia panamericana, chihuahua
chihuahua 31200
MX   31200
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key16855971
MDR Text Key314414210
Report Number9680794-2023-00317
Device Sequence Number1
Product Code NFK
Combination Product (y/n)N
Reporter Country CodeSF
PMA/PMN Number
K020430
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCAR-02800
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NOT REPORTED.; NOT REPORTED.
-
-