BAXTER HEALTHCARE CORPORATION MINICAP TRANSFER SET; SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
|
Back to Search Results |
|
Catalog Number 5C4482 |
Device Problems
Fluid/Blood Leak (1250); Defective Component (2292)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: initial reporter address: (b)(6).The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the twist clamp of a minicap transfer set leaked when the clamp was closed; the "inner place" of clamp was abnormal.This occurred during use of the device for peritoneal dialysis therapy.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was received for evaluation with a minicap attached to the female connector.A visual inspection with the naked eye noted a broken occluder leg.The reported condition was verified.The cause of the leak was due to the broken occlude leg.The cause of the damage could not be determined; however, exposure to chemical agents as hydrogen peroxide, alcohol or bleach as listed on product packaging can damage the transfer set materials.A nonconformance has been opened to address this issue.A batch review was conducted and there were no deviations found related to this condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|