SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC DISPOSABLE CIRCUITS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
|
Back to Search Results |
|
Model Number 100/905/340-CH |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/01/2021 |
Event Type
malfunction
|
Event Description
|
It was reported that during the pre-use check, the customer noticed the relief pressure value did not go up to the assigned one.It was suspected there was leakage from the air circuit.No patient injury was reported.
|
|
Manufacturer Narrative
|
Other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).One sample was received for investigation.Visual inspection confirmed that in the middle of the circuit, it confirmed that there was a scratch on it about 50 cm from the tip of the connector on the device side.No other analysis was performed.The root cause was unable to be determined.A corrective and preventative action was opened by the supplier to address the reported issue.
|
|