• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC DISPOSABLE CIRCUITS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC DISPOSABLE CIRCUITS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number 100/905/340-CH
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2021
Event Type  malfunction  
Event Description
It was reported that during the pre-use check, the customer noticed the relief pressure value did not go up to the assigned one.It was suspected there was leakage from the air circuit.No patient injury was reported.
 
Manufacturer Narrative
Other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).One sample was received for investigation.Visual inspection confirmed that in the middle of the circuit, it confirmed that there was a scratch on it about 50 cm from the tip of the connector on the device side.No other analysis was performed.The root cause was unable to be determined.A corrective and preventative action was opened by the supplier to address the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC DISPOSABLE CIRCUITS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16861931
MDR Text Key314479386
Report Number3012307300-2023-04983
Device Sequence Number1
Product Code BTL
UDI-Device Identifier15019517160215
UDI-Public15019517160215
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/18/2023
Device Model Number100/905/340-CH
Device Catalogue Number100/905/340-CH
Device Lot Number200518
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/27/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-