• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PRISMATIK DENTALCRAFT, INC. COMFORT HARD-SOFT SPLINT; THERMOFORM MOUTHGAURD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PRISMATIK DENTALCRAFT, INC. COMFORT HARD-SOFT SPLINT; THERMOFORM MOUTHGAURD Back to Search Results
Lot Number 1120156939
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Type  malfunction  
Event Description
It was reported that the patient had a reaction to the comfort hard soft splint that was issued.It is unclear when the patient received the device, when the patient first used the device, or when the reaction occurred or resolved.However, it is noted that the patient experienced irritation, metallic taste inside lip and around gum tissue.The irritation was not visible but can feeling.The device was worn for a few nights (exact dates unknown) and the symptoms went away after a few nights (exact dates unknown).The patient has an allergy to latex.Update: patient questionnaire received on (b)(6) 2023 with additional information.The device was delivered on (b)(6) 2023 and used that same evening.There was soft tissue irritation and a metallic taste in the mouth.It is unknown when the device was discontinued, but reaction resolved within 24hrs.There was no medical treatment required.There is an allergy to latex.With regard to the device: the provider was advised that a label will be sent to return appliance for further research.The device was cleaned using an ultrasonic device with denture cleaner.The patient rinsed with toothpaste and a brush.
 
Manufacturer Narrative
The device has been returned.Once the investigation is complete a supplemental report will be submitted.This complaint will be kept on record for track and trending purposes.
 
Manufacturer Narrative
Capa 2023-006 the device was returned, the investigation has been completed and the results are as follows: dhr results: the dhr was reviewed and there was no evidence discovered to indicate that a product defect or non-conformity contributed to the issue.Supplier (erkodent) reviewed the associated material lot and confirmed no manufacturing deviation or abnormality.An allergic reaction is possible.Additionally, erkodent reported no further complaints for this material lot.(please see attached coa) lot#ekd5.0-11481 was manufactured from september 14, 2020 and was assigned an expiration of september 2023.Stock product reviewed results: no stock product was available for review since the device was fabricated per physician's prescription only.Investigation method/results: complaint investigator reviewed the returned device.The results were summarized and the pictures were attached.(see below) roughness - the flange was smooth.Internal/external surfaces were smooth.Crack - no major crack was found.Delamination - layers were intact and did not separate.Discoloration - the device appeared clear and transparent with yellowish tint.General cleanliness - the returned device appeared to be clean, free of debris or foreign particles.The returned device was visually inspected and no defect or abnormality was observed.There was no evidence found to indicate that the reported issue was caused by the device itself.Root cause: a root cause for this complaint cannot be explicitly determined.It is possible that reactions could be caused by mouthwash, toothpaste, or soaking material.However, the customer did not provide the information regarding how the patient handled and maintained the device.Per the reported information, the patient has an allergy to latex.Ifu 9091 rev 5.0 (comfort h/s bite splint) contains the following statement in the warning section: "use only clear, cool water to wash the device.Do not clean or soak in mouthwash.Do not use denture cleanser.Do not use hot water (for cleaning).Do not use alcohol or hydrogen peroxide.Do not place in direct sunlight.Keep away from heat sources." ifu 9091 rev 5.0 (comfort h/s bite splint) contains the following statement for the cleaning procedures in the general safety and precautions section: "brush and floss your teeth before use.Rinse mouth well with clean water before inserting the device.If patient uses mouthwash, all traces of mouthwash should be removed by thoroughly rinsing out mouth with water.Rinse bite splint well with clean, cool water before and after use.Clean bite splint with clean, cool water only and let air dry.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMFORT HARD-SOFT SPLINT
Type of Device
THERMOFORM MOUTHGAURD
Manufacturer (Section D)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer (Section G)
PRISMATIK DENTALCRAFT, INC.
2144 michelson drive
irvine CA 92612
Manufacturer Contact
herbert crane
2144 michelson drive
irvine, CA 92612
9783130850
MDR Report Key16863227
MDR Text Key314495688
Report Number3011649314-2023-00253
Device Sequence Number1
Product Code MQC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K121365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 10/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/03/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Lot Number1120156939
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/13/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/29/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/27/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age67 YR
Patient SexFemale
Patient Weight54 KG
Patient RaceWhite
-
-