• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW ORTHOPAEDICS AG BICON-PLUS TITANIUM SHELL 5-56 NON-CEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW ORTHOPAEDICS AG BICON-PLUS TITANIUM SHELL 5-56 NON-CEM; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS Back to Search Results
Catalog Number 75003742
Device Problems Loss of Osseointegration (2408); Migration (4003)
Patient Problems Inadequate Osseointegration (2646); Joint Laxity (4526); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/04/2023
Event Type  Injury  
Manufacturer Narrative
H10: internal complaint reference: (b)(4).
 
Event Description
It was reported that, after a thr surgery that had been performed on (b)(6)2008, the patient experienced a mobilization of the acetabular cup.This adverse event was solved via revision surgery performed on (b)(6)2023, in which a bicon-plus titanium shell 5-56 non-cem, a metal metal ball head 28s cone 12/14, and bicon-plus insert were explanted.The current health status of the patient is unknown.
 
Manufacturer Narrative
Additional information: h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: it was reported that, after a total hip replacement surgery, the patient experienced a mobilization of the acetabular cup.This adverse event was solved via revision surgery, in which a bicon-plus titanium shell 5-56 non-cem, a metal metal ball head 28s cone 12/14, and bicon-plus insert were explanted.The current health status of the patient is unknown.The device used in treatment was not returned for investigation.A product evaluation was not possible.A review of the production documentation did not detect any deviation that could have contributed to the reported failure mode.The review of historical complaints for the alleged device revealed no additional similar complaints reported for the same batch, and one additional similar complaints for the same product number over the past 12 months with similar failure mode.A review of past corrective actions was performed.No further escalation is required.A review of the risk management documentation verifies the failure mode, occurrence and severity of the reported issue.The instructions for use (lit.No.12.23, ed 03/21) states "osseointegration problem/loosening of implant not related to bone-ingrowth" as a ¿potential medical device problems¿ resulting from a hip arthroplasty.No clinically sufficient data was provided to perform a medical investigation.The performed investigation does not lead to an accurately determined cause.There is no evidence that the reported devices failed to meet manufacturing specifications upon release for distribution.Due to insufficient information, it is not possible to speculate about factors which could have contributed to the reported event.There is no need for further actions because of the limited information provided.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.Corrected data: b1 (type of event), h6 (health effect - clinical code, medical device problem code).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BICON-PLUS TITANIUM SHELL 5-56 NON-CEM
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, POROUS
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
Manufacturer (Section G)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ   CH-5000
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16869458
MDR Text Key314566228
Report Number9613369-2023-00143
Device Sequence Number1
Product Code MBL
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 06/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/04/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/07/2015
Device Catalogue Number75003742
Device Lot NumberD0804032
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/31/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/08/2008
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
BICON-PLUS PE INSERT ANTIL 5/28 NON-CEM, LOT#:0411; METAL METAL BALL HEAD 28S CONE 12/14, LOT#:C080126
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age85 YR
-
-