SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE)
|
Back to Search Results |
|
Catalog Number C49291349-NLJ |
Device Problem
Device Damaged Prior to Use (2284)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 07/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Visual inspection showed that returned sample has the connector correctly installed.Based on the inspection, it can't be confirmed failure mode reported as loose connector.Root cause cannot be addressing since the failure mode reported is not confirmed.No corrective actions were taken since the investigation did not reveal the failure reported.
|
|
Event Description
|
It was reported that the connection on the patient side of the connector was loose.This was found during pre-test.No patient injury was reported.
|
|