Catalog Number 397002-001 |
Device Problems
Battery Problem (2885); Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/24/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The driver will be returned to syncardia for evaluation.The results will be provided in a follow-up mdr.
|
|
Event Description
|
The customer, a syncardia certified hospital, reported the companion 2 driver received an emergency battery error.
|
|
Manufacturer Narrative
|
Alarm history and patient data file review found review found new alarms recorded in the companion 2 driver's data file: on apr 24th there was an emergency battery error alarm that occurred multiple times.Visual inspection of external components revealed no abnormalities.Visual inspection of internal components revealed a damaged capacitor on the power management board though unrelated to this complaint.This finding handled per standard non-conformance procedures.Companion 2 driver passed all incoming functional testing.The battery was then subjected to a ten-minute battery discharge test.During this testing the battery performed as intended.Failure investigation for this complaint confirmed the reported issue via alarm history data review.The complaint was not replicated via functional/observational testing.The root cause of the customer reported emergency battery alarm could not be conclusively determined.Failure investigation identified no test failure, damage, or abnormalities that could have contributed to the customer reported alarm.Driver functioned as designed.Patient was switched to a back-up driver with no reported adverse impact.This issue weill be monitored and trended as part of the customer complaint process.Syncardia has completed its evaluation and is closing this file.(if new or additional information is received in the future, syncardia will file a follow-up mdr.) (b)(4).Follow-up report 1.
|
|
Search Alerts/Recalls
|