MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION EVOLUT FX DCS; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV Back to Search Results
Model Number D-EVOLUTFX-2329
Device Problem Failure to Advance (2524)
Patient Problems Stroke/CVA (1770); Muscle Weakness (1967); Visual Impairment (2138); Rupture (2208); Speech Disorder (4415)
Event Date 04/24/2023
Event Type  Injury  
Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the attempted implant of this transcatheter bioprosthetic valve, the delivery catheter system (dcs) was unable to advance through the aortic arch.A piece of calcium was observed and it appeared to stop the dcs from advancing.After multiple unsuccessful attempts to advance, the dcs was removed from the patient.A non-medtronic valve was implanted.One day post-operation, it was reported that the patient had a small arch rupture and signs of stroke.No treatment was reported.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Updated b5.Updated h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received which reported that the physician felt the delivery catheter system (dcs) could have played a role in the rupture, but it was difficult to predict.Thoracic endovascular aortic repair (tevar) was performed as a result of the arch rupture.It was noted that the calcified aortic arch played a role in the rupture.The stroke was confirmed via magnetic resonance imaging (mri).The patient experienced weakness in their arms and legs, slurred speech and impaired vision.The physician's opinion is that the dcs could have played a part in the stroke, based on the dcs getting caught in that section of arch and after multiple attempts, it could have moved around calcium debris.Unspecified treatment was provided for the stroke.No additional adverse patient effects were reported.
 
Manufacturer Narrative
Updated data: b5.G2.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information was received that the treatment the patient received was a follow up and monitored by the neurologist.It was noted the patient regained normal vision and speech.No adverse patient effects were reported.
 
Manufacturer Narrative
Conclusion: difficulties advancing the delivery catheter system (dcs) through the access vessel is known to be related to factors such as patient anatomy and physician technique, including guidewire and introducer sheath selection.In this case, it was noted that there was calcium preventing the advancement of the dcs.This indicates that the probable cause of the advancement difficulties was patient anatomy, but this cannot be confirmed with the limited information available.Vascular related complications, such as rupture, are a known potential adverse patient effect per the device instructions for use (ifu), and are typically related to patient factors (anatomy, comorbidities, etc.), and/or procedural effects (sheath used, user technique, puncture cut location, etc.).The physician felt the dcs could have played a role in the rupture, but it was difficult to predict.There was no information to suggest a device malfunction or a failure to meet manufacturing specifications was related to these events.Multiple factors can influence the onset of a stroke.Its etiology may include embolization of calcific debris, patient medical history, and procedural factors.¿stroke (ischemic or haemorrhagic), transient ischemic attack (tia), or other neurological deficits¿ are listed as potential adverse events associated with the use of the valve in the device ifu.The physician's opinion is that the dcs could have played a part in the stroke, based on the dcs getting caught in that section of arch and after multiple attempts, it could have moved around calcium debris.Updated: h6.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOLUT FX DCS
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIV
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16881535
MDR Text Key314711055
Report Number2025587-2023-01957
Device Sequence Number1
Product Code NPT
UDI-Device Identifier00763000365677
UDI-Public00763000365677
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date 07/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberD-EVOLUTFX-2329
Device Catalogue NumberD-EVOLUTFX-2329
Device Lot Number0011630243
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2023
Date Device Manufactured02/09/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Disability; Life Threatening;
Patient Age81 YR
Patient SexFemale
-
-