Catalog Number 50-1010
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001032347-2023-00158.0001032347-2023-00160.D10: item# 50-1010; lot# 133340.Item# 50-1010; lot# 665630.G2: foreign - event occurred in japan.Customer has indicated that the product will not be returned to zimmer biomet for investigation, as the product was discarded.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
It was reported that during inspection at the receiving warehouse, there was debris found in the sterile packaging.No customer or patient was involved.Attempts have been made and no further information has been provided.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.No product was returned.A visual inspection was conducted on the customer provided pictures.The photos show debris inside of the sterile packaging.Reported event did not occur in an operating room or as part of a medical procedure; medical records are not available for review.Review of the device history records identified no deviations or anomalies during manufacturing.The likely condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the provided work instructions during manufacturing.If any further information which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY POWERDRIVER P2
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer (Section G)
BIOMET MICROFIXATION
1520 tradeport drive
jacksonville FL 32218
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key16884929
MDR Text Key314729345
Report Number0001032347-2023-00159
Device Sequence Number1
Product Code GEY
UDI-Device Identifier00841036063504
UDI-Public(01)00841036063504(17)240509(10)133340
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number50-1010
Device Lot Number133340
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/09/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberNI
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-