• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST PAUL CADD CASSETTE RESERVOIRS - FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST PAUL CADD CASSETTE RESERVOIRS - FLOW STOP; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7301-24
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/20/2021
Event Type  malfunction  
Event Description
It was reported that during use of disposable the pump exhibited no cassette alarm.No patient injury was reported.
 
Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No problems or issues were identified during the device history record review.Two samples were received without their original packaging and inside a plastic bag.Visual inspection was performed at 12 to 16 inches under normal conditions of illumination.The sample did not present any damage, scuffs, pinch marks, cracks, crazing, cuts, etc.Could cause the failure mode reported.During the functional testing, the sample was fully primed and connected without difficulty, the pump was set running and no alarms were activated.The complaint was not confirmed no root cause was determined due complaint was not confirmed.The actions taken were not performed due complaint was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD CASSETTE RESERVOIRS - FLOW STOP
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16887754
MDR Text Key314778156
Report Number3012307300-2023-05306
Device Sequence Number1
Product Code FPA
UDI-Device Identifier10610586027215
UDI-Public10610586027215
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K040636
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number21-7301-24
Device Catalogue Number21-7301-24
Device Lot Number4114178
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2021
Was the Report Sent to FDA? No
Date Manufacturer Received05/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-