• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDINAL HEALTH KANG Y-SITE EXT SET ENFIT; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDINAL HEALTH KANG Y-SITE EXT SET ENFIT; GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS Back to Search Results
Model Number 8884705008E
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
An investigation is currently underway.Upon completion, the results will be forwarded.
 
Event Description
The customer reported leaking at proximal end of connect site; broken sets.
 
Manufacturer Narrative
A device history record review could not be performed because a lot number was not received with the complaint.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.Three samples were returned for evaluation.Visual and functional inspections were performed, and no failures were found.Based on the information available to us, we were unable to confirm the event.In the meantime, all information received will be used for further tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANG Y-SITE EXT SET ENFIT
Type of Device
GASTROINTESTINAL TUBES WITH ENTERAL SPECIFIC CONNECTORS
Manufacturer (Section D)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer (Section G)
CARDINAL HEALTH
777 west street
mansfield MA 02048
Manufacturer Contact
jill saraiva
777 west street
mansfield, MA 02048
5086183640
MDR Report Key16888604
MDR Text Key314856493
Report Number1282497-2023-10755
Device Sequence Number1
Product Code PIF
UDI-Device Identifier30884521657612
UDI-Public30884521657612
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number8884705008E
Device Catalogue Number8884705008E
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-