OAKDALE LEVEL 1 FAST FLOW FLUID WARMER; WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION Back to Search Results
Catalog Number 8002910
Device Problems Display or Visual Feedback Problem (1184); Temperature Problem (3022)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Manufacturer Narrative
B3: date of event and d4: udi number is unknown.No information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr, 803.56, when additional reportable information becomes available.
 
Event Description
It was reported, that the device was not heating consistently.And display lights were not staying on.Patient involvement unknown.
 
Manufacturer Narrative
Other text: d10.Device available for evaluation; h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was received with no cuffs, an obsolete air detector, old style float switch that was cracked, ground stud on back panel had been removed and a screw put in its place, fluid ingression on printed circuit board (pcb), cracked return tube, and non original equipment manufacturer (oem) parts used on heaters and pump.Functional testing confirmed the complaint as the device had no flow rate and was not heating due to pump not operating.The operational light emitting diode (led) was not illuminating.The root cause was incorrect installation of pump and fluid ingression on printed circuit board (pcb) from cracked return tube due to user interface.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.No action taken due to the condition of the device.It was deemed beyond economical repair and was scrapped.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 FAST FLOW FLUID WARMER
Type of Device
WARMER, BLOOD, NON-ELECTROMAGNETIC RADIATION
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16891561
MDR Text Key314824886
Report Number3012307300-2023-05352
Device Sequence Number1
Product Code BSB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
BK020043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8002910
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/12/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-