MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME Back to Search Results
Model Number N/A
Device Problems Failure to Cut (2587); Pressure Problem (3012); Patient Device Interaction Problem (4001)
Patient Problem Laceration(s) (1946)
Event Date 04/13/2023
Event Type  Injury  
Manufacturer Narrative
This event is recorded with zimmer biomet under (b)(6).This medwatch is being filed as an initial report.Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported a dermatome caused patient harm, it was noted it skipped across the skin causing lacerations and did not glide as supposed to.Further it wasa noted that there seemed to be a problem in regard to air pressure and the way it glided over an aubergine.Follow up is in process and to date no further informaiton is available.
 
Manufacturer Narrative
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Review of the most recent repair record could not be completed because the device was not returned for evaluation and repair.Lot/serial identification is necessary for review of device history records, and lot/serial identification was not provided.A definitive root cause cannot be determined.The event cannot be confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No additional event information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key16894642
MDR Text Key314832766
Report Number0001526350-2023-00447
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/09/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100100
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received10/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
ITEM:00885100206LOT:22/270283
Patient Outcome(s) Other;
Patient SexPrefer Not To Disclose
-
-