|
Model Number H700494 |
Device Problem
High impedance (1291)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/05/2023 |
Event Type
malfunction
|
Event Description
|
During an atrial fibrillation ablation procedure, impedance issues resulted in the procedure being cancelled.It was noted that there was high impedance (above 150 ohms) during ablation.The rf cables and catheter were exchanged, but the issue persisted and the procedure was cancelled.There are no reported adverse patient consequences.
|
|
Manufacturer Narrative
|
One ampere¿ rf ablation generator was received for evaluation.The product functioned as anticipated during testing with no abnormalities identified that would result in the reported complaint.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.No non-conformances associated with the reported event were identified.Based on the information provided to abbott and the investigation performed, root cause of the field reported event could not be conclusively determined as the returned ampere generator functioned as anticipated, with no anomalies identified that would result in the reported complaint.
|
|
Search Alerts/Recalls
|
|
|