• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1¿, 63 CM LENGTH, 8.5F; INTRODUCER, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1¿, 63 CM LENGTH, 8.5F; INTRODUCER, CATHETER Back to Search Results
Model Number 407453
Device Problem Material Puncture/Hole (1504)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/13/2023
Event Type  malfunction  
Event Description
During an atrial fibrillation ablation procedure, an attempt was made to insert the device for a transseptal puncture, and the needle pierced the dilator.The physician is not certain if it was due to the tortuosity of the vessel and the needle was not reshaped.The device was exchanged, and the procedure was completed with no adverse patient health consequences.
 
Manufacturer Narrative
One 8.5f swartz braided introducer dilator was received for evaluation.Visual inspection revealed the dilator was perforated 1.48¿ proximal to the distal tip.There were also multiple bends throughout the dilator.A needle/stylet assembly from current inventory was inserted and advanced through the dilator; however, the brk needle exited through the aforementioned perforation.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported dilator puncture could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SWARTZ¿ BRAIDED TRANSSEPTAL GUIDING INTRODUCER, SL1¿, 63 CM LENGTH, 8.5F
Type of Device
INTRODUCER, CATHETER
Manufacturer (Section D)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
ST. JUDE MEDICAL
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key16900491
MDR Text Key314893216
Report Number3005334138-2023-00191
Device Sequence Number1
Product Code DYB
UDI-Device Identifier05414734205771
UDI-Public05414734205771
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052644
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number407453
Device Catalogue Number407453
Device Lot Number8873167
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/28/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/02/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age50 YR
Patient SexMale
Patient Weight151 KG
-
-