• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS LUCEA 40; DEVICE, MEDICAL EXAMINATION, AC POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS LUCEA 40; DEVICE, MEDICAL EXAMINATION, AC POWERED Back to Search Results
Model Number ARD568601998
Device Problems Fracture (1260); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/24/2023
Event Type  malfunction  
Event Description
On 24th april, 2023 getinge became aware of an issue with one of examination equipment - lca 40.It was stated the light handle was broken, based on photographic evidence the cover was broken resulting in missing particles.There was no injury reported, however we decided to report the issue in abundance of caution as any particles falling off during examination may lead to potential infection of the patient.
 
Manufacturer Narrative
Additional information will be provided following the conclusion of the investigation.H3 other text : device not returned to manufacturer.
 
Event Description
Manufacturer's reference number (b)(4).
 
Manufacturer Narrative
According to the information provided by getinge technician, the affected part was replaced (ard368605998 - handle interface with fork - lucea 40).Based on the information collected, it was established that when the event occurred, the examination light did not meet its specification, since cracks in covers, resulting in missing plastic particles could be considered as technical deficiency, and in this way the device contributed to the event.The provided information does not indicate if upon the event occurrence, the device was or was not being used for patient treatment.When reviewing reportable events for this type of issue we were able to establish that the received incidents are occurring at a moderate ratio.We have been also able to confirm that the issue investigated herein has never led to serious injury or worse.Root cause analysis was performed by subject matter experts at manufacturing site.As they stated, the most likely root cause of this malfunction is a violent rotation of the light head.There was re11-052 test performed which proved that the stop in the lighthead resists to 50 strong rotations and this is compliant to the standard ul60601-1.We believe the related devices are performing correctly in the market.We also believe that if the manufacturer¿s recommendation had been followed the incident could have been avoided.Getinge shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUCEA 40
Type of Device
DEVICE, MEDICAL EXAMINATION, AC POWERED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key16902514
MDR Text Key314909869
Report Number9710055-2023-00347
Device Sequence Number1
Product Code KZF
Combination Product (y/n)N
Reporter Country CodeSN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARD568601998
Device Catalogue NumberARD568601998
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/04/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/26/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-