It was reported that during a cryo ablation procedure, the dilator would not lock into the sheath.The sheath was replaced which resolved the issue.It was then reported that the st segment started increasing gradually.Immediately after the first ablation, the st segment increased greatly, resulting in bradycardia.The ablation was discontinued and air embolism was detected in the left atrium (la) near the vena cava, as confirmed by angiography and echocardiography.The air in the blood vessel was suctioned using a suction catheter which resolved the issue. the case was completed with cryo.Post operatively, an mri was performed and no head issues were observed.No further patient complications have been reported as a result of this event.
|
Mapping catheter 2ach20.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.If information is provided in the future, a supplemental report will be issued.
|