During an atrial fibrillation procedure, communication issues resulted in procedural delays.The ampere connect cable was replaced and the issue was resolved.The procedure was completed with no adverse consequences to the patient.It was reported that the catheter looked distorted on ensite x, and the distal egm was not displaying accurately.Attempted to exchange several catheters, the tactisys, the rf cable, the ampere, and power cycle with no resolution.The ampere connect cable was exchanged and the issue was resolved.The procedure was continued with no adverse patient consequences.
|