• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SCHILLER AG TEMPUS LS-MANUAL; LOW ENERGY DEFIBRILLATOR Back to Search Results
Model Number 00-3020
Device Problems Loose or Intermittent Connection (1371); Unable to Obtain Readings (1516)
Patient Problem Cardiac Arrest (1762)
Event Date 03/22/2023
Event Type  Injury  
Event Description
It has been reported to philips that while working a cardiac arrest, the monitor and ls were properly attached with pads and 3 lead.No rhythm was seen on either the monitor or ls.After a change in pads, there was still no rhythm seen on either screen.A user report was received related to a reported product problem which is currently being investigated.Further updates will be provided when the investigation is completed.
 
Manufacturer Narrative
This report is based on information provided by a philips remote service engineer, schiller (equipment manufacturer) and a rdt technical investigations engineer (remote diagnostic technologies) and has been investigated by the philips complaint handling team.Philips received a complaint on the tempus ls and tempus pro indicating that no rhythm was observed on either device.This complaint is concerning investigation into the tempus ls.While the device was noted to be in use, there was reportedly no patient or user harm or impact.The reported complaint and good faith effort response from the customer states that pads and 3 lead were connected and that with a change to new good pads still no rhythm was observed.Diagnostics determined the device would require return for investigation and the tempus ls was received at schiller where a technical investigation was completed.The device was tested and found to be functioning satisfactorily.The device worked as intended.The returned device was removed from service and processed following local procedures.A technical investigation was completed at rdt to review the log and rescue files for impedance results.The event described by the customer was observed as the impedance measured by the pads was less less than 25 o, logs also indicate unplug and plug of pads (as user described as change of pads), the impedance measured was still less than 25 o even after a change of pads.The tempus ls displays "check defibrillator electrodes".This message is reported when the impedance range falls outside the acceptable range per the manual (25 o-250 o).When the impedance measurement is out of range (less than 25o and or greater than 250o), tempus ls does not display the ecg signals.Based on the information available and testing conducted, the cause of the reported problem is unknown, the device functioned satisfactorily and no problem was found with the device.The reported problem was not confirmed.Based on the information available, further action is necessary at this time.While no harm was alleged, recurrence of this failure mode during clinical use could cause or contribute to death or serious injury.Therefore, this failure mode meets the definition of a reportable malfunction.Appropriate regulatory reporting actions have been taken.The data entered in this complaint record will be utilized for product quality and safety improvements per the post market surveillance and risk management processes.The customer was provided a replacement device to resolve the issue.It has been concluded that no further action is required at this time.If additional information is received the complaint file will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TEMPUS LS-MANUAL
Type of Device
LOW ENERGY DEFIBRILLATOR
Manufacturer (Section D)
SCHILLER AG
altgasse 68
baar 6341
SZ  6341
Manufacturer (Section G)
SCHILLER AG
altgasse 68
baar 6341
SZ   6341
Manufacturer Contact
tanya deschmidt
ascent 1, aerospace centre
aerospace boulevard
farnborough GU14 -6XW
UK   GU14 6XW
MDR Report Key16918306
MDR Text Key315072250
Report Number3003832357-2023-00299
Device Sequence Number1
Product Code LDD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K200849
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number00-3020
Device Catalogue Number989706001681
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/17/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-