• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER GMBH TARGET DEVICE GAMMA3®; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER GMBH TARGET DEVICE GAMMA3®; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 1320-0111
Device Problem Failure to Align (2522)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/15/2023
Event Type  malfunction  
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event Description
As reported: "gamma drill is hitting the nail when going through the sleeve on the jig.This is the jig from the gamma3 basic set.They removed the wire and drilled without doing it over the wire to successfully complete procedure." it was additionally reported that there was a 30 minute surgical delay.
 
Manufacturer Narrative
Please note the corrections to d3 and d9/h3.The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device disposition unknown.
 
Event Description
As reported: "gamma drill is hitting the nail when going through the sleeve on the jig.This is the jig from the gamma3 basic set.They removed the wire and drilled without doing it over the wire to successfully complete procedure." it was additionally reported that there was a 30 minute surgical delay.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TARGET DEVICE GAMMA3®
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16919463
MDR Text Key315094915
Report Number0009610622-2023-00151
Device Sequence Number1
Product Code HSB
UDI-Device Identifier04546540716774
UDI-Public04546540716774
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K123401
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1320-0111
Device Catalogue Number13200111
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-