• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 STATLOCK SL FOLEY, UNKNOWN CODE; UNKNOWN STATLOCK SL FOLEY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 STATLOCK SL FOLEY, UNKNOWN CODE; UNKNOWN STATLOCK SL FOLEY Back to Search Results
Device Problem Difficult to Open or Close (2921)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/21/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that the statlock foley component attached to the catheter could not be opened.
 
Manufacturer Narrative
The reported event is confirmed - cause unknown.Visual inspection noted one (1) statlock with no packaging.Statlock clamp would not open, and tab was bent from force of trying to open clamp, but would not pop open clamp.Therefore, product does not meet specifications.The potential root cause could be process error.Unable to perform a dhr review since the lot number was not provided.A labeling review is not required because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the statlock foley component attached to the catheter could not be opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STATLOCK SL FOLEY, UNKNOWN CODE
Type of Device
UNKNOWN STATLOCK SL FOLEY
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key16919488
MDR Text Key315097847
Report Number1018233-2023-03371
Device Sequence Number1
Product Code EYJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/04/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/12/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/02/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-