C.R. BARD, INC. (COVINGTON) -1018233 STATLOCK SL FOLEY, UNKNOWN CODE; UNKNOWN STATLOCK SL FOLEY
|
Back to Search Results |
|
Device Problem
Difficult to Open or Close (2921)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that the statlock foley component attached to the catheter could not be opened.
|
|
Manufacturer Narrative
|
The reported event is confirmed - cause unknown.Visual inspection noted one (1) statlock with no packaging.Statlock clamp would not open, and tab was bent from force of trying to open clamp, but would not pop open clamp.Therefore, product does not meet specifications.The potential root cause could be process error.Unable to perform a dhr review since the lot number was not provided.A labeling review is not required because labeling could not have prevented the reported failure.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
|
|
Event Description
|
It was reported that the statlock foley component attached to the catheter could not be opened.
|
|