MEDTRONIC SOFAMOR DANEK USA, INC CD HORIZON® BALLAST¿ SPINAL SYSTEM; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
|
Back to Search Results |
|
Model Number 25740018590 |
Device Problem
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
Patient Problem
Inadequate Osseointegration (2646)
|
Event Type
Injury
|
Event Description
|
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient who had l5/s fusion f ailure.It was reported that patient had s1 screw removal s2ai screw replacement due to s1 screw loosening s2ai set screw dislocation.Patient had l5/s fusion failure due to screw loosening and dislocation.There was no patient symptom reported after revision surgery.There were no further complications reported regarding the event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
H3: product analysis #(b)(4) lot# ca19l080 visual and microscopic inspection confirmed the set screw has some deformation on the node of the screws.There is some smearing/damaged to the nodes and the screw saddles from what appears to be the rods moving between the saddles and set screw.The threads of the set screws have been damaged.Functionally tested with a bone screw and confirmed the screw would still thread into the head of the bone screw.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|