PERFUSION SYSTEMS EOPA 3D VENTED ARTERIAL CANNULA; CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO Back to Search Results
Model Number 78220
Device Problem Product Quality Problem (1506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2020
Event Type  malfunction  
Event Description
Medtronic received information that during use this eopa arterial cannula had an air bubble in the wall of the cannula.The cannula was used to complete the case.There is no adverse patient effect associated with this event.
 
Manufacturer Narrative
Device evaluation summary: visual inspection shows a bubble in the body of the returned cannula approximately 0.020 inches in length.Reason for return was confirmed.Conclusion: after investigation at medtronic, the complaint is confirmed for microbubbles in the cannula body.Microbubbles occur during the manufacturing process; a certain size of fully encapsulated bubbles are acceptable per the manufacturing specifications and do not affect the form or function of the device.Review of the device history record found no abnormalities during manufacturing that would cause or contribute to the reported event.Complaints received from december 2019 through january 2021 for similar model numbers were reviewed and showed no trends warranting escalation related to this occurrence.There were no adverse patient effects as a result of this occurrence.Medtronic has made its supplier aware of this occurrence and will continue to monitor for future occurrences and trends.This regulatory report is being submitted as part of a retrospective review and remediation per d00953163 as part of a capa action.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EOPA 3D VENTED ARTERIAL CANNULA
Type of Device
CATHETER, CANNULA AND TUBING, VASCULAR, CARDIO
Manufacturer (Section D)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer (Section G)
PERFUSION SYSTEMS
7611 northland dr
brooklyn park MN 55428
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key16929281
MDR Text Key315686290
Report Number2184009-2023-00264
Device Sequence Number1
Product Code DWF
UDI-Device Identifier00643169486249
UDI-Public00643169486249
Combination Product (y/n)N
Reporter Country CodeTH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Expiration Date11/30/2021
Device Model Number78220
Device Catalogue Number78220
Device Lot Number2018111620
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
Patient Weight57 KG
-
-