COVIDIEN LP LLC NORTH HAVEN ENDO GIA; STAPLE, IMPLANTABLE Back to Search Results
Model Number EGIA60AMT
Device Problems Entrapment of Device (1212); Retraction Problem (1536); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Unspecified Tissue Injury (4559)
Event Date 04/27/2023
Event Type  malfunction  
Event Description
According to the reporter, during the laparoscopic low rectum resection, when starting to fire on the rectum with a manual handle and 60mm purple reload, the surgeon heard a first crack sound.Continued the firing process and heard a second crack sound to be able to reach the end point.The jaws were locked on tissue and it was then impossible retract the knife using normal technique.With 2 instruments (forceps) to have leverage action on the jaws was a success.Previous attempt with a hook to pull back the knife was unsuccessful.A partial tear in the rectum was noted.The surgical incision was extended up to 12-13cm and there was a minimal tissue damage.When using another manual handle with the same lot number and a 45mm purple reload, when starting the firing process, the stapler did not fire.Squeezing the trigger had no effect at all.The surgeon was able to press the green button.A powered stapler and a new 45mm purple reload were used to complete the case.The surgeon performed drainage as a precaution which resulted to extension of surgical time for 1 hour and 10 minutes.
 
Manufacturer Narrative
D10 concomitant product/s: egiaustnd egiaustnd endogia ultra univ std stap lot #:p3b0002 egiaustnd egiaustnd endogia ultra univ std stap lot #:p3b0002 egia45amt egia45amt egia 45 artic med thick sulu lot #:n2m0444y medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the reload was fully fired and the jaws of the reload were clamped.The knife bar assembly was advanced to the distal end.It was reported that the jaws of the device remain closed on tissue and after firing the device, there was resistance while attempting to retract the knife.The reported issues were confirmed.The most likely cause was determined to be manufacturing related.Bent knife laminates can occur from improper component orientation.The bent knife laminates push the interlock legs down causing the interlock to fail.This issue can also occur when the device is applied on over-indicated tissue.Internal process improvements have been initiated to mitigate this issue.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO GIA
Type of Device
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer (Section G)
COVIDIEN LP LLC NORTH HAVEN
195 mcdermott rd
north haven CT 06473
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key16930368
MDR Text Key315200180
Report Number1219930-2023-01949
Device Sequence Number1
Product Code GDW
UDI-Device Identifier10884523003215
UDI-Public10884523003215
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K111825
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 07/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/15/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEGIA60AMT
Device Catalogue NumberEGIA60AMT
Device Lot NumberN2J0206Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/03/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/11/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age48 YR
Patient SexFemale
Patient Weight65 KG
-
-