MAUDE Adverse Event Report: COOK INC BAKRI TAMPONADE BALLOON CATHETER; OQY INTRAUTERINE BALLOON

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC BAKRI TAMPONADE BALLOON CATHETER; OQY INTRAUTERINE BALLOON Back to Search Results
Model Number G30673
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2023
Event Type  malfunction  
Event Description
As reported, the stopcock handle of 'bakri tamponade balloon catheter' detached from the device during a uterine balloon tamponade.The user attempted to reattach it but was unsuccessful and the device was removed from the patient.The procedure was completed by using another j-sos-100500 from an unknown lot.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Additional information regarding event details, patient anatomy and outcome has been requested but is not available at this time.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown or unavailable.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Correction: h6 (annex c and annex d).Correction: h10.Gouges were visible in the handle, likely from unintended stress by the user.Based on the information provided, inspection of the returned device, and the results of the investigation, cook has concluded unintended user error contributed to this incident in the form of excessive force applied to the stopcock.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Event summary as reported, the stopcock handle of 'bakri tamponade balloon catheter' detached from the device during a uterine balloon tamponade.The user attempted to reattach it but was unsuccessful and the device was removed from the patient.The procedure was completed by using another j-sos-100500 from an unknown lot.A section of the device did not remain inside the patient¿s body.The patient did not require any additional intervention due to this occurrence.The patient did not experience any adverse effects due to this occurrence.Investigation ¿ evaluation a document-based investigation was performed including a review of complaint history, device history record (dhr), instructions for use (ifu), and quality control procedures, as well as visual inspection and functional test of the device were conducted.The complaint device was returned for evaluation.The blue portion of the stopcock was separated from the clear portion.A review of the dhr for the reported complaint device lot revealed no recorded non-conformances relevant to the failure mode.A database search did not identify any other events associated with the reported device lot.Based on the available information, cook has concluded that there is no evidence suggesting nonconforming product exists either in house or in the field.A review of relevant manufacturing documents was conducted.It was concluded that the device aspect in question was functionally inspected by quality control and no notable gaps in production or processing controls were noted.There is no indication that a design or process related failure mode contributed to the reported event.Current controls for manufacturing are in place to assure functionality and device integrity prior to shipping.Cook also reviewed product labeling.The product ifu provides the following information to the user related to the reported failure mode: how supplied "upon removal from the package, inspect the product to ensure no damage has occurred." based on the information provided, inspection of the returned device, and the results of the investigation, the most probable cause of the reported event could not be determined from the available information.Cook will continue monitoring of similar complaints and has notified the appropriate personnel of this event.¿ per a review of risk documentation, no further action is required.¿ this report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BAKRI TAMPONADE BALLOON CATHETER
Type of Device
OQY INTRAUTERINE BALLOON
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer (Section G)
COOK INC
750 daniels way
bloomington IN 47404
Manufacturer Contact
jason crouch
750 daniels way
bloomington, IN 47404
8123392235
MDR Report Key16939962
MDR Text Key315371273
Report Number1820334-2023-00609
Device Sequence Number1
Product Code OQY
UDI-Device Identifier10827002306735
UDI-Public(01)10827002306735(17)241220(10)14430605
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K170622
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 09/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberG30673
Device Catalogue NumberJ-SOS-100500
Device Lot Number14430605
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient SexFemale
-
-