Model Number M3002A |
Device Problem
Device Alarm System (1012)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/14/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the alarm is not triggered.It is unknown if the device was in clinical use at the time of the event, no adverse event or patient harm was reported.
|
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.Section e reporting institution phone #: (b)(6).
|
|
Manufacturer Narrative
|
Philips received a complaint on the intellivue multi measurement server x2 indicating that the alarm was not triggered.The field service engineer (fse) went onsite and found that the alarm volume was set to 0.The fse set the alarm volume to 3 to resolve the issue.The device was confirmed to be operating per specifications.Based on the information available and the testing conducted, the cause of the reported problem was the monitor alarm volume settings.The reported problem was confirmed.
|
|