PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH INTELLIVUE MULTI MEASUREMENT SERVER X2 Back to Search Results
Model Number M3002A
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/14/2023
Event Type  malfunction  
Event Description
It was reported that the alarm is not triggered.It is unknown if the device was in clinical use at the time of the event, no adverse event or patient harm was reported.
 
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.Section e reporting institution phone #: (b)(6).
 
Manufacturer Narrative
Philips received a complaint on the intellivue multi measurement server x2 indicating that the alarm was not triggered.The field service engineer (fse) went onsite and found that the alarm volume was set to 0.The fse set the alarm volume to 3 to resolve the issue.The device was confirmed to be operating per specifications.Based on the information available and the testing conducted, the cause of the reported problem was the monitor alarm volume settings.The reported problem was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIVUE MULTI MEASUREMENT SERVER X2
Type of Device
INTELLIVUE MULTI MEASUREMENT SERVER X2
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key16941557
MDR Text Key315370534
Report Number9610816-2023-00234
Device Sequence Number1
Product Code DSI
UDI-Device Identifier00884838000261
UDI-Public00884838000261
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K150310
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM3002A
Device Catalogue NumberM3002A
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received06/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-