• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER Back to Search Results
Model Number 87047
Device Problem Material Perforation (2205)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2023
Event Type  malfunction  
Manufacturer Narrative
The intellanav mifi oi catheter was returned to boston scientific for analysis.The returned device passed all tests performed, showing no evidence of defects that could have contributed to the reported event, however, a cut was observed in the catheter shaft.The complaint investigation conclusion code is manufacturing deficiency.
 
Event Description
During a catheter ablation procedure to treat ventricular tachycardia in the inguinal region a intellanav mifi open-irrigated ablation catheter was selected for use.It was reported that "the circuit was not recognized" by the generator on electrodes 1 and 2 and a unknown magnetic tracking error message was observed.The catheter cable was first replaced, however, the event did not improve.The catheter was exchanged and the issue was resolved.The procedure was completed with no patient complications.However, device analysis revealed a cut in the catheter shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
dc a330
saint paul, MN 55112
6515826168
MDR Report Key16942464
MDR Text Key315364692
Report Number2124215-2023-23979
Device Sequence Number1
Product Code OAD
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 05/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/16/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/21/2023
Device Model Number87047
Device Catalogue Number87047
Device Lot Number0027354770
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/22/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/18/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-