BOSTON SCIENTIFIC CORPORATION INTELLANAV MIFI OPEN-IRRIGATED ABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER
|
Back to Search Results |
|
Model Number 87047 |
Device Problem
Material Perforation (2205)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The intellanav mifi oi catheter was returned to boston scientific for analysis.The returned device passed all tests performed, showing no evidence of defects that could have contributed to the reported event, however, a cut was observed in the catheter shaft.The complaint investigation conclusion code is manufacturing deficiency.
|
|
Event Description
|
During a catheter ablation procedure to treat ventricular tachycardia in the inguinal region a intellanav mifi open-irrigated ablation catheter was selected for use.It was reported that "the circuit was not recognized" by the generator on electrodes 1 and 2 and a unknown magnetic tracking error message was observed.The catheter cable was first replaced, however, the event did not improve.The catheter was exchanged and the issue was resolved.The procedure was completed with no patient complications.However, device analysis revealed a cut in the catheter shaft.
|
|