• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; SIMILAR DEVICE D133601, PMA # P030031/S053

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER; SIMILAR DEVICE D133601, PMA # P030031/S053 Back to Search Results
Catalog Number D133605IL
Device Problems Tear, Rip or Hole in Device Packaging (2385); Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/19/2023
Event Type  malfunction  
Manufacturer Narrative
The product has not returned for analysis, however, a picture of the complaint issue was provided by the customer.Evaluation is still in progress.The analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Note: the ¿suspected medical device¿ reported in this report is not marketed in usa or approved by the fda.However, it is being reported as biosense webster considers this as a similar device to thermocool® smart touch¿ electrophysiology catheter approved under p030031/s053.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#: (b)(4).
 
Event Description
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch¿ electrophysiology catheter and the sterilized pouch where the device is located was torn compromising sterility.It was reported that before the operation, the sterilization package was broken after the carton was opened.A second device was used to complete the operation.There was no adverse event reported on patient.Device was not used in patient.
 
Manufacturer Narrative
It was reported that a patient underwent a cardiac ablation procedure with a thermocool® smart touch¿ electrophysiology catheter and the sterilized pouch where the device is located was torn compromising sterility.It was reported that before the operation, the sterilization package was broken after the carton was opened.A second device was used to complete the operation.There was no adverse event reported on patient.Device was not used in patient.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection of the returned device was performed following bwi procedures.Pictures were also provided by the customer to aid in the product investigation.According to pictures provided by the customer, a hole was observed on the pouch of the device.A visual analysis in the lab of the returned product also revealed damage in the packaging; which could compromise the sterility of the catheter.All units are inspected prior leaving the facility as there are functional tests and inspections at control points based on the process flow diagram (pfd) per its part number to avoid this type of damage from leaving the facility.The damage reported, could be related to the process of storing, packaging or shipping of the device, however, this cannot be conclusively determined.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Manufacturer Narrative
On 24-may-2023, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH¿ ELECTROPHYSIOLOGY CATHETER
Type of Device
SIMILAR DEVICE D133601, PMA # P030031/S053
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key16946956
MDR Text Key315489670
Report Number2029046-2023-01054
Device Sequence Number1
Product Code LPB
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 07/10/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/23/2023
Device Catalogue NumberD133605IL
Device Lot Number30950292M
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Date Manufacturer Received06/15/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/24/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
THMCL SMARTTOUCH,TC,F,C3,OBL.
-
-