Model Number 4674
Device Problems Positioning Problem (3009); High Capture Threshold (3266)
Patient Problem Vascular Dissection (3160)
Event Date 05/08/2023
Event Type  Injury  
Event Description
It was reported that this brady lead was attempted but unsuccessful due to unknown product performance issue.Additional information was obtained, which indicates that there's a placement difficulty due to anatomy and dissection.The lead also has a high threshold and no pacing options.The device was never in service.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that this brady lead was attempted but unsuccessful due to unknown product performance issue.Additional information was obtained, which indicates that there's a placement difficulty due to anatomy and dissection.The lead also has a high threshold and no pacing options.The device was never in service.No additional adverse patient effects were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Analysis found no evidence of device defect, malfunction or damage outside the bounds of normal medical use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY X4 SPIRAL S
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16947219
MDR Text Key315396188
Report Number2124215-2023-24600
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526524554
UDI-Public00802526524554
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S398
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/08/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number4674
Device Catalogue Number4674
Device Lot Number895105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Hospitalization; Required Intervention;
Patient Age80 YR
Patient SexMale
-
-