Model Number 4674 |
Device Problems
Positioning Problem (3009); High Capture Threshold (3266)
|
Patient Problem
Vascular Dissection (3160)
|
Event Date 05/08/2023 |
Event Type
Injury
|
Event Description
|
It was reported that this brady lead was attempted but unsuccessful due to unknown product performance issue.Additional information was obtained, which indicates that there's a placement difficulty due to anatomy and dissection.The lead also has a high threshold and no pacing options.The device was never in service.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that this brady lead was attempted but unsuccessful due to unknown product performance issue.Additional information was obtained, which indicates that there's a placement difficulty due to anatomy and dissection.The lead also has a high threshold and no pacing options.The device was never in service.No additional adverse patient effects were reported.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Analysis found no evidence of device defect, malfunction or damage outside the bounds of normal medical use.
|
|