SYNTHES GMBH SYNPOR SQUARE SHEET-STERILE 50MMX50MM/0.8MM THICK; PLATE, CRANIOPLASTY, PREFORMED, ALTERABLE Back to Search Results
Catalog Number 08.510.120S
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/23/2023
Event Type  malfunction  
Event Description
Device report from depuy synthes reports an event in china as follows: it was reported that on (b)(6) 2023, during the surgery, opened the packaging and some particles fall off.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.This report is for one (1) synpor square sheet-sterile 50mmx50mm/0.8mm thick.This is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: d9: complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Investigation summary: the photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device from attachment.Visual analysis of the photo revealed that the synpor sheet 50*50 t0.8 was found to be delaminated across several sections of the sheet and there are small flakes at the surface of the table.The fragments appears to be broken across the mid section.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for the synpor sheet 50*50 t0.8.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: part# 08.510.120s lot # ds7007282 manufacturing site: werk selzach logistik release to warehouse date: 01 may 2021 supplier: dsm biomedical expiration date: 01 april 2026 a manufacturing record evaluation was performed for the finished article lot and no non-conformances were identified.Note: dhr information was retrieved from pi-16796777185346658 as it is the same lot number.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: the device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.D2: additional procodes: ftm, ftl.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the synpor sheet 50*50 t0.8 was found to be delaminated across several sections of the sheet and there are small flakes at the surface of the table.The fragments appears to be broken across the mid section.A dimensional inspection was not performed as it is not applicable to the complaint condition.The overall complaint was confirmed as the observed condition of the synpor sheet 50*50 t0.8 would contribute to the complained device issue.Based on the investigation findings, the root cause is traced to design.The product issue has been addressed through depuy synthes quality system.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.If information is obtained that was not available for this medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNPOR SQUARE SHEET-STERILE 50MMX50MM/0.8MM THICK
Type of Device
PLATE, CRANIOPLASTY, PREFORMED, ALTERABLE
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
WERK SELZACH
bohnackerweg 5
selzach CO 2545
SZ   2545
Manufacturer Contact
kate karberg
eimattstrasse 3
oberdorf 4436
SZ   4436
3035526892
MDR Report Key16948099
MDR Text Key315408067
Report Number8030965-2023-06260
Device Sequence Number1
Product Code GWO
UDI-Device Identifier07611819315853
UDI-Public(01)07611819315853
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K051879
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number08.510.120S
Device Lot NumberDS7007282
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/07/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-