Model Number G148
Device Problems Delayed Charge Time (2586); Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/05/2023
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) had reached a battery status indicator of explant.The battery status indicator four months previous indicated ten months remaining to explant.The explant indicator was found to be due to extended charge times.Technical services discussed, and device replacement was scheduled.Currently, evidence suggests the device remains in service.No adverse patient effects were reported.
 
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) had reached a battery status indicator of explant.The battery status indicator four months previous indicated ten months remaining to explant.The explant indicator was found to be due to extended charge times.Technical services discussed, and device replacement was scheduled.Currently, evidence suggests the device remains in service.No adverse patient effects were reported.To date, no additional information has been received.Should additional follow-up information be provided in the future, an updated report will be issued.
 
Manufacturer Narrative
Evidence suggests this device remains implanted; therefore, technical analysis cannot be conducted.Without a returned device it is not possible to definitively confirm how the device may have contributed to the complaint incident.If pertinent information is provided in the future, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CRT-D
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16949108
MDR Text Key315421862
Report Number2124215-2023-24662
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/02/2016
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number103407
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/12/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial