HEARTWARE, INC. HEARTWARE VENTRICULAR ASSIST SYSTEM - DRIVELINE; VENTRICULAR (ASSISST) BYPASS
|
Back to Search Results |
|
Model Number 1104 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the driveline exhibited an unspecified damage, and a repair was performed.The driveline remains in use.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Investigation of this event is pending and a supplemental report will be sent upon its completion.Additional information has been requested regarding the details of the event, but it was not available at the time of this report.If additional information is received, the event will be updated and a supplemental report will be sent.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A supplemental is being submitted for additional information.Product event summary: the driveline cable associated with (b)(6) was not returned for evaluation.There was no evidence that (b)(6) had previously been serviced.Review of the manufacturing documentation confirmed that the associated device met all requirements for release.The reported event could not be confirmed due to insufficient evidence.Based on the risk documentation and historical review of similar events, a possible root cause of the reported damage to the driveline may be attributed, but not limited, to factors such as normal wear over time, improper handling, design issues, and/or exposure to uv light.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|