MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM TIB BASEPLATE - CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM TIB BASEPLATE - CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5520-B-300
Device Problem Unintended Movement (3026)
Patient Problem Bone Fracture(s) (1870)
Event Date 04/25/2023
Event Type  Injury  
Manufacturer Narrative
Reported event: an event regarding periprosthetic fracture and subsidence involving a triathlon baseplate was reported.The event was not confirmed.Method & results: -product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: it was reported that the patient was revised due to tibial fracture and baseplate subsidence.The exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event Description
It was reported that the patient's right knee was revised after patient complaint of pain following 2 falls.Pre-operatively, surgeon reported tibial fracture and baseplate subsidence.The baseplate, insert, and femoral component were revised.Rep confirmed femur was revised to convert to a ps knee, and that there were no allegations against the revised insert.The patellar component was retained.Rep confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PRIM TIB BASEPLATE - CEMENTED
Type of Device
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
marisol santiago
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16951359
MDR Text Key315441876
Report Number0002249697-2023-00549
Device Sequence Number1
Product Code MBH
UDI-Device Identifier07613327050318
UDI-Public07613327050318
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 05/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/30/2023
Device Model Number5520-B-300
Device Catalogue Number5520-B-300
Device Lot NumberBLH4TB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/25/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/21/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient Age74 YR
Patient SexFemale
Patient Weight83 KG
-
-