During the device evaluation, the nozzle was observed to be deformed.Due to wear of angle wire, the bending angle in up direction did not meet the standard value.Due to wear of angle wire, the play of up/down knob was out of the standard value.The adhesive on the bending cover (a-rubber) was peeling and chipped.The connecting tube, camera cover (c-cover), and distal end were scratched.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 16 years since the subject device was manufactured.Based on the results of the investigation, the customer¿s reported issue could not be confirmed.The foreign material could not be identified, but there is a possibility that the foreign matter remained due to physical damage to the nozzle and deviation of the reprocess from the instruction manual.The event can be detected/prevented by following the instructions: the inspection method for the event is described as follows in the instruction manual (operation edition) "chapter 3 preparation and inspection 3.2 inspection of the endoscope and 3.6 inspection of the combination function with related equipment".[endoscope inspection] 8.Visually check that there are no abnormalities such as abnormal protrusion, dents, deformation, etc.In the air and water feed nozzles at the tip of the endoscope.[inspection of the cleaning function of the objective lens surface] 1.When the button is pressed while blocking the hole of the air and water supply button with your finger, check that water flows throughout the endoscope image.Olympus will continue to monitor field performance for this device.
|