LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 16-02-85 |
Device Problem
Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/21/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
There was no patient involvement.Livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in hudson, florida.A livanova field service representative was dispatched to the facility to investigate the device and could not confirm the reported issue, since the device was emptied by the customer.Level sensor shafts were cleaned by livanova technician in case these were sticking preventing the customer from knowing when the patient tank was full.Missing bracket for the vacuum canister was installed.Unit was filled with water and tested: all tests passed showing no signs of leaks.The unit was returned to service.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
Livanova deutschland received a report that a heater-cooler system 3t device was leaking water from front and back, during cleaning operations.There was no patient involvement.
|
|
Manufacturer Narrative
|
H10: based on available information, it cannot be ruled out that leak was caused by the user who inadvertently overfilled the unit during cleaning operation.
|
|