Catalog Number CDS0702-XTW |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 04/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
This is filed to report a break in the clip introducer.It was reported a mitraclip procedure was to be performed.A mitraclip xtw was selected for treatment.While outside the patient and while the clip introducer was flushed through the clip introducer port, a break in the clip introducer occurred.Therefore, the mitraclip xtw was not used and was replaced.There was no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The reported broken clip introducer flush port was confirmed during returned device analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported clip introducer flush port break is due to environmental stress cracking (mother nature/environment), and is investigated in an exception.Corrective action will be implemented in the future to address this issue by replacing the current luer with a new luer.Abbott will continue to trend the performance of these devices.
|
|