• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD); MITRAL VALVE REPAIR DEVICES Back to Search Results
Catalog Number CDS0702-XTW
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report a break in the clip introducer.It was reported a mitraclip procedure was to be performed.A mitraclip xtw was selected for treatment.While outside the patient and while the clip introducer was flushed through the clip introducer port, a break in the clip introducer occurred.Therefore, the mitraclip xtw was not used and was replaced.There was no clinically significant delay in the procedure.No additional information was provided.
 
Manufacturer Narrative
The reported broken clip introducer flush port was confirmed during returned device analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported clip introducer flush port break is due to environmental stress cracking (mother nature/environment), and is investigated in an exception.Corrective action will be implemented in the future to address this issue by replacing the current luer with a new luer.Abbott will continue to trend the performance of these devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MITRACLIP G4 CLIP DELIVERY SYSTEM (MDD)
Type of Device
MITRAL VALVE REPAIR DEVICES
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16964756
MDR Text Key316046455
Report Number2135147-2023-02218
Device Sequence Number1
Product Code NKM
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P100009
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/30/2023
Device Catalogue NumberCDS0702-XTW
Device Lot Number21031R1038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/12/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/31/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-