The customer reported a complaint that "the thermogard ivtm system sn (b)(4) cabel measured incorrectly, it's not cooling and makes loud noise" which was not confirmed during the event log review and during the functional testing.However, a wrong cable used on the thermogard ivtm system was observed.A monitor cable was connected instead of a patient cable which caused to show incorrect temperature measurement, likely attributed to user error.Unrelated to the reported complaint, worn-out guide rollers on the pump rotor were noted upon visual inspection.The observed cosmetic damages root cause is likely normal wear and tear.The thermogard xp ivtm system was manufactured in (b)(6) 2014 and is 9 years old, past beyond its service life of 5 years.The white guide rollers were replaced to address the observed cosmetic damage.The event log review indicated no discrepancy.The thermogard system passed the functional testing without errors.Following service, the thermogard xp ivtm system passed the final functional without any issues.
|
The thermogard ivtm system sn (b)(4) cabel measured incorrectly, it's not cooling and makes loud noise.It is unknown where the problem occurred.However, patient use information was requested, but no additional information was provided.
|