It was reported that during a rotator cuff revision surgery, when the user tried to tap the bone using a healicoil dilator to insert a healicoil regenesorb anchor, the first implant titanium anchor was inadvertently found.This resulted in the breakage of the tip of the healicoil dilator into the titanium anchor, from which it could not be removed.The procedure was completed without surgical delay using a back-up device.No further complications were reported.No further information is available.
|
H3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.Based on the information available, there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A clinical review states that the patient impact beyond the reported possible retained particles could not be definitively determined.Micro-motion or movement cannot be predicted and there is potential risk for tissue inflammation and/or pain.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|