• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH¿ NORMAL SALINE SYRINGE; SALINE, VASCULAR ACCESS FLUSH Back to Search Results
Model Number 306546
Device Problem Component Missing (2306)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/27/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation summary: a device history record review was completed by our quality engineer team for provided material number 306546 and lot number 3010631.The review did not reveal any detected abnormalities during the production process that could have contributed to this defect and all quality tests were found to be within specification.As a sample was unavailable for return, a thorough sample investigation could not be completed.Based on the investigation results, an exact cause for this incident could not be identified.
 
Event Description
It was reported that the bd posiflush¿ normal saline syringe flush was missing cap on syringe luer end, still sealed in cellophane wrapper.The following information was provided by the initial reporter: "flush was missing cap on syringe luer end, still sealed in cellophane wrapper." additional info received on (b)(6) 2023.The date of the finding was one day prior to reporting to bd product complaints department, only one syringe was found and reported, it is unknown if other syringes were found and not reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH¿ NORMAL SALINE SYRINGE
Type of Device
SALINE, VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer (Section G)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
Manufacturer Contact
phillip emmert
9450 south state street
sandy, UT 84070
8448235433
MDR Report Key16970477
MDR Text Key316109804
Report Number1911916-2023-00331
Device Sequence Number1
Product Code NGT
UDI-Device Identifier30382903065463
UDI-Public(01)30382903065463
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161552
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number306546
Device Catalogue Number306546
Device Lot Number3010631
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-