Model Number 71440005
Device Problems Fracture (1260); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/26/2023
Event Type  malfunction  
Event Description
It was reported that during a tka surgery while impacting the femoral component the tip of the femoral implant impactor broke inside the patient.The surgical site was washed out and pieces were retrieved with a suction tip.The procedure was resumed, without any delay, using the same device.The patient was not harmed as a consequence of this problem.
 
Manufacturer Narrative
Complaint reference number: (b)(4).
 
Manufacturer Narrative
H3, h6.The associated device was returned and evaluated.The visual inspection revealed one of the cam arm covers is missing from the device.The cam arm cover was not returned.The device shows signs of significant wear and use.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history based on the historical data revealed a similar event for the listed batch, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that this event was previously identified and addressed; a stress relief process was established for acetal bumpers, samples were tested in order to include the stress relief operation in the manufacturing flow and to include the inspection of stress relief temperature chart to determine compliance.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL IMPLANT IMPACTOR
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key16970545
MDR Text Key315697700
Report Number1020279-2023-01084
Device Sequence Number1
Product Code JDG
UDI-Device Identifier00885556034156
UDI-Public00885556034156
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/29/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71440005
Device Catalogue Number71440005
Device Lot Number13MM06881
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/10/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-