• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DRÄGERWERK AG & CO. KGAA PERSEUS A500; ANESTHESIA UNITS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DRÄGERWERK AG & CO. KGAA PERSEUS A500; ANESTHESIA UNITS Back to Search Results
Catalog Number MK06000
Device Problems Gas Output Problem (1266); Failure to Analyze Signal (1539); Failure to Deliver (2338); No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/20/2023
Event Type  malfunction  
Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.H3 other text : on-going.
 
Event Description
It was reported the unit went into boot cycle.No patient injury reported.
 
Manufacturer Narrative
The replaced parts were available for investigation.Non of the parts showed signs of burn or increased heat development.When tested in a lab device the pba blower a500 showed at least one faulty hall sensor.During a test run of the pba ther.Contr.Unit m16.2 in a lab device in total 5 reboots of less than 15 seconds were observed, but no cycling reboot.No hardware failures could be found.Although no specific hardware failure could be identified, the therapy control unit m16.2 can be regarded as cause for the (cycling) reboot.The subsequent blower failure was caused by a faulty hall sensor signal.The power supply (m7.3) was found to be fully functional.In case of a reset of the processors of the therapy control unit m16 therapy is discontinued for a maximum of 15 seconds, before therapy will be resumed with the latest valid settings.If for any reasons therapy was not restarted within 15 seconds an acoustical alarm will be given.Manual ventilation remains possible.
 
Event Description
It was reported the unit went into boot cycle.No patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERSEUS A500
Type of Device
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck
GM 
Manufacturer (Section G)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM   23542
Manufacturer Contact
moislinger allee 53-55
lübeck 23542
4518822868
MDR Report Key16972102
MDR Text Key315865640
Report Number9611500-2023-00187
Device Sequence Number1
Product Code BSZ
UDI-Device Identifier04048675253600
UDI-Public(01)04048675253600(11)211015(17)220222(93)MK06000-40
Combination Product (y/n)N
PMA/PMN Number
K133886
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMK06000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/24/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-