Catalog Number SEPX-8-6-40-135 |
Device Problem
Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/08/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Physician attempted to use a protégé rx self expanding stent with a 5mm spider fx embolic protection device during patient treatment.The ifu (instruction for use) was followed during preparation, procedure, post-procedure.There was no damage noted to packaging (i.E.Shelf carton, hoop/tray).There were no issues noted when removing the device from the hoop/tray.The vessel was not pre-dilated but was post-dilated.The device did not pass through a previously-deployed stent.No resistance was encountered when advancing the device.After the spider was in place, when pushed the carotid artery stent, it was found that the guide wire at the tip of the spider was broken.At this time, the stent had just been half deployed, and the operator terminated the surgery for the safety of the patient.After the spider and stent were taken out of the body together, it was found that the guide wire of the umbrella had slipped and was not completely broken.However, it could not continue to be used after a doctor's evaluation.No intervention required for the removal of the devices and were safely removed from the patient.No vascular damage visible on x-ray.A replacement spider and stent were used to complete procedure successfully.No patient injury reported.
|
|
Manufacturer Narrative
|
Product analysis the device was returned with the touhy-borst tightened.The device was returned with the stent deployed, the stent was confirmed as 40mm, the stent retainer and inner were examined and no damage noted medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|