Model Number HL 20
Device Problem Failure to Pump (1502)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/16/2023
Event Type  malfunction  
Event Description
It was reported that during start up the hl 20 pump displayed the error message "beltslip".No harm to any person was reported.Complaint id: (b)(4).
 
Manufacturer Narrative
The investigation is ongoing.A getinge technician will investigate the cardiohelp.A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
Complaint id: (b)(4).
 
Manufacturer Narrative
It was reported that a single pump displayed the error message "beltslip".The failure occurred during start up.A getinge field service technician (fst) was sent for investigation and repair on 2023-06-01.The tacho board was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The tacho board was investigated by the getinge life cycle engineering on 2023-08-10 with the following outcome.The reported failure could not be replicated during the investigation.Therefore no exact root cause could be determined.The most probable root cause was identified as an electronic defect at the two optical sensors or their amplifiers.The device was manufactured on 2022-03-18.The device history record (dhr) of the hl 20 was reviewed on 2023-06-07.There is no indication that manufacturing issues occurred, thus production related influences are unlikely to have contributed to the reported failure.Based on the results the reported failure "error message "beltslip" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
JULIA KAPFENBERGER
neue rottenburger strasse 37
hechingen
Manufacturer Contact
neue rottenburger strasse 37
hechingen 
MDR Report Key16977614
MDR Text Key315703009
Report Number8010762-2023-00230
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K943803
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL 20
Device Catalogue Number701034386
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/10/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-