MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WALLFLEX ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00516920
Device Problems Use of Device Problem (1670); Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2023
Event Type  Injury  
Event Description
It was reported to boston scientific corporation on (b)(6) 2023, that a wallflex esophageal partially covered stent was to be implanted to treat an esophageal stenosis during an esophageal stenting procedure performed on april 29, 2023.During the procedure, the tip detached when the device was advanced through the esophagus.The device was removed from the patient and the procedure was not completed.There were no patient complications as a result of this event.A photo of the device provided by the complainant shows that the stent was partially deployed, and the tip was detached.
 
Manufacturer Narrative
Block h6: imdrf device code a0501 captures the reportable event of tip detached.Imdrf device code a15 captures the reportable event of stent partially deployed.
 
Manufacturer Narrative
Blocks b5, d7, e1, e2, e3, h6 (impact codes and device codes) and h8 have been updated with the additional information received on may 23, 2023, may 24, 2023, and may 29, 2023.Imdrf device code a0501 captures the reportable event of tip detached.Imdrf device code a15 captures the reportable event of stent partially deployed.
 
Event Description
It was reported to boston scientific corporation on april 29, 2023, that a wallflex esophageal partially covered stent was to be implanted to treat an esophageal stenosis during an esophageal stenting procedure performed on (b)(6) 2023.During the procedure, the tip detached when the device was advanced through the esophagus.The device was removed from the patient and the procedure was not completed.There were no patient complications as a result of this event.A photo of the device provided by the complainant shows that the stent was partially deployed, and the tip was detached.Additional information received on may 23, 2023, may 24, 2023, and may 29, 2023.It was reported that the wallflex esophageal partially covered stent was to be implanted to treat a benign esophageal stenosis.During the procedure, the tip detached inside the patient's upper esophageal sphincter.The tip was removed together with the sg dilator metal guidewire.The patient's anatomy was not tortuous and was not dilated prior to stent placement.Note: it was reported that the stent was attempted to be implanted to treat a benign esophageal stricture.However, per the wallflex esophageal partially covered stent system directions for use, the stent is indicated for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.The stent is not indicated to be placed for a benign esophageal stricture.
 
Event Description
It was reported to boston scientific corporation on (b)(6) 2023, that a wallflex esophageal partially covered stent was to be implanted to treat an esophageal stenosis during an esophageal stenting procedure performed on (b)(6) 2023.During the procedure, the tip detached when the device was advanced through the esophagus.The device was removed from the patient and the procedure was not completed.There were no patient complications as a result of this event.A photo of the device provided by the complainant shows that the stent was partially deployed, and the tip was detached.Additional information received on may 23, 2023, may 24, 2023, and may 29, 2023: it was reported that the wallflex esophageal partially covered stent was to be implanted to treat a benign esophageal stenosis.During the procedure, the tip detached inside the patient's upper esophageal sphincter.The tip was removed together with the sg dilator metal guidewire.The patient's anatomy was not tortuous and was not dilated prior to stent placement.Note: it was reported that the stent was attempted to be implanted to treat a benign esophageal stricture.However, per the wallflex esophageal partially covered stent system directions for use, the stent is indicated for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas.The stent is not indicated to be placed for a benign esophageal stricture.
 
Manufacturer Narrative
Block h6: imdrf device code a0501 captures the reportable event of tip detached.Imdrf device code a15 captures the reportable event of stent partially deployed.Block h10: a wallflex esophageal stent and delivery system were received for analysis.Visual examination of the returned device found the stent fully deployed and expanded.The tip was detached and was not returned.The delivery system was also kinked in two sections and the outer sheath was buckled.Microscopic inspection was performed, and the inner sheath showed evidence of adhesive.A media inspection was performed of a photo provided by the complainant, and it was observed that the stent was partially deployed, and the tip was detached.No other problems were noted to the stent and delivery system.The reported events of stent partially deployed, and tip detached were confirmed.A labeling review was performed and from the information available, this device was used in a manner inconsistent per the instructions for use (ifu)/product label.The ifu states, "the wallflex esophageal partially covered stent system is indicated for maintaining esophageal luminal patency in esophageal strictures caused by intrinsic and/or extrinsic malignant tumors, and occlusion of concurrent esophageal fistulas." the stent is not indicated to be placed for a benign esophageal stricture.Taking all available information into consideration, the investigation concluded that the reported events and observed failures were likely due to factors encountered during the procedure.It may be that lesion characteristics, how the device was handled, and/or the technique used by the physician (force applied), limited the performance of the device and contributed to the reported events and observed failures.Therefore, a review and analysis of the all the available information indicated that the most probable cause is adverse event related to procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WALLFLEX ESOPHAGEAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16984404
MDR Text Key315760496
Report Number3005099803-2023-02630
Device Sequence Number1
Product Code ESW
UDI-Device Identifier08714729765264
UDI-Public08714729765264
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K073266
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 10/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/20/2023
Device Model NumberM00516920
Device Catalogue Number56453
Device Lot Number0027347666
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/02/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/20/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexMale
-
-