Model Number 25335
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/01/2023
Event Type  malfunction  
Event Description
It was reported that a foreign object was observed.An encore 26 inflation device was selected for use.However, a white object was noted floating inside the device.The issue was observed outside the patient and the procedure was completed with another of the same device.No complications were reported.
 
Manufacturer Narrative
Device evaluated by mfr.: the encore 26 device was returned for analysis.Analysis of the returned product revealed that no foreign materials were observed, consequently the reported complaint cannot be confirmed.
 
Event Description
It was reported that a foreign object was observed.An encore 26 inflation device was selected for use.However, a white object was noted floating inside the device.The issue was observed outside the patient and the procedure was completed with another of the same device.No complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENCORE 26
Type of Device
SYRINGE, BALLOON INFLATION
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
model farm road
cork
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key16985176
MDR Text Key316212001
Report Number2124215-2023-25524
Device Sequence Number1
Product Code MAV
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K955869
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/23/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number25335
Device Catalogue Number25335
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/21/2023
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-