SMITH & NEPHEW, INC. JOURNEY FEM IMPL IMPACTOR; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Model Number 74012831 |
Device Problem
Solder Joint Fracture (2324)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/02/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complaint reference number: (b)(4).
|
|
Event Description
|
It was reported that, during a tka surgery, a journey fem impl impactor snapped at the neck junction close to the bumper.No pieces fell inside the patient.The procedure was resumed, without any delay, using a s+n back-up device.No injury was reported due to this issue.
|
|
Manufacturer Narrative
|
Section h10: the device was not returned for evaluation.However, the photographs were reviewed and revealed that the device fractured at the weld of the handle.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Internal complaint reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|
|