• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CLEARSTREAM TECHNOLOGIES LTD. SLEEK OTW PTA CATHETER; PTA BALLOON DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CLEARSTREAM TECHNOLOGIES LTD. SLEEK OTW PTA CATHETER; PTA BALLOON DILATATION CATHETER Back to Search Results
Model Number 426-3010X
Device Problems Deflation Problem (1149); Difficult to Remove (1528)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2023
Event Type  malfunction  
Manufacturer Narrative
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device pending return.
 
Event Description
It was reported that during a percutaneous transluminal angioplasty procedure, the balloon catheter allegedly would not deflate.It was further reported that the balloon was allegedly not possible to be pulled out of the unknown catheter sheath introducer.Reportedly, the balloon and catheter sheath introducer were removed together.There was no reported patient injury.
 
Event Description
It was reported that during a percutaneous transluminal angioplasty procedure, the balloon catheter allegedly would not deflate.It was further reported that the balloon was allegedly not possible to be pulled out of the unknown catheter sheath introducer.Reportedly, the balloon and catheter sheath introducer were removed together.There was no reported patient injury.
 
Manufacturer Narrative
H10: manufacturing review: a complaint history review was performed.This is the first complaint reported for this product/lot number combination.Therefore, a device history record review is not required.Investigation summary: the device was not returned for evaluation.The result of the investigation is inconclusive for the reported device deflation and removal issues.The root cause for the reported device deflation and removal issues could not be determined based upon the available information.Labeling review: the instructions for use for the sleek otw product was reviewed and contains the following information relevant to the reported event: balloon characteristics ¿ please check the package label for the rated burst pressure (rbp).It is important that the balloon not be inflated beyond the rated burst pressure.Pressures in excess of rated burst pressure may cause the balloon to burst.Compliance charts are provided with each product.Please note that balloon diameters may vary within manufacturing tolerances.All inflations should be viewed under fluoroscopy.The sleek® otw balloons reach their nominal diameter at 6 atm (608 kpa).Precautions ¿ if resistance is felt upon removal, then the balloon, guidewire and the sheath/guide catheter should be removed together as a unit, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and sheath/guide catheter as a unit and withdrawing both together, using a gently twisting motion combined with traction.¿ before removing catheter from sheath/guide catheter it is very important that the balloon is completely deflated.¿ proper functioning of the catheter depends on its integrity.Care should be used when handling the catheter.Damage may result from kinking, stretching, or forceful wiping of the catheter.Procedure: insertion and inflation procedure: note: a 0.014¿ (0.356 mm) guidewire must be inserted in the sleek® otw catheter across the balloon during any inflation of the balloon.Make sure that the balloon sleeve has been removed from the catheter balloon.Enter the vessel percutaneously using the standard seldinger technique over the appropriate guidewire for the size catheter being used.Advance the catheter across the lesion with fluoroscopic guidance using accepted percutaneous transluminal angioplasty technique and inflate the balloon to the appropriate pressure.Note: do not inflate the balloon or advance the catheter unless the guidewire is in place.Deflation and withdrawal ¿ deflate the balloon by drawing a vacuum with a 20 ml or larger syringe.Note: the larger the syringe diameter, the greater the suction that is applied.For maximum deflation a 50 ml syringe is recommended.¿ gently withdraw the catheter.As the balloon exits the vessel, use a smooth, gentle, steady, counterclockwise motion.If resistance is felt upon removal then the balloon and the sheath should be removed together as a unit under fluoroscopic guidance, particularly if balloon rupture or leakage is known or suspected.This may be accomplished by firmly grasping the balloon catheter and sheath as a unit and withdrawing both together, using a gentle twisting motion combined with traction.¿ apply pressure to the insertion site according to standard practice or hospital protocol for percutaneous vascular procedures.H10: g3, h6 (method) h11: h6 (result, conclusion) h11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SLEEK OTW PTA CATHETER
Type of Device
PTA BALLOON DILATATION CATHETER
Manufacturer (Section D)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI  N A
Manufacturer (Section G)
CLEARSTREAM TECHNOLOGIES LTD.
moyne upper
enniscorthy, co. wexford N A
EI   N A
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key16990721
MDR Text Key315823695
Report Number9616666-2023-00087
Device Sequence Number1
Product Code DQY
UDI-Device Identifier00801741133756
UDI-Public(01)00801741133756
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number426-3010X
Device Catalogue Number426-3010X
Device Lot NumberCMFZ0010
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/12/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-